Pressure calibration service

 

Pressure calibration service

The calibration laboratory for “absolute pressure and positive overpressure” allows the calibration of analogue and digital pressure measuring devices and pressure sensors with an electrical output signal (e. g. 0...10 V, 4...20 mA, ...).

Calibration is carried out as a comparison measurement against a dead weight tester utilised as the reference standard. The dead weight tester is used to pressurise the instrument under test with a defined pressure from 1...1200 bar. Depending on the actual measuring range, the smallest possible measuring uncertainties are 0.7 mbar.

 

Sample calibration certificate

Sample DAkkS certificate

 

Please get in touch with us if you have any other questions about our calibration service:

Service Center
Tel.: +49 5605 803-0
Fax: +49 5605 803-555
This email address is being protected from spambots. You need JavaScript enabled to view it.


For fast handling of your calibration and maintenance orders, please use our product return procedure.